Main -> Heating units, refrigerating units -> Temperature controllers -> Incubators, cooled incubators |
Cooled Incubators / Storage Cooled Incubators, Memmert |
Series ICP compressor cooled with TwinDISPLAY The generation of compressor-cooled incubators with the unique ControlCOCKPIT for quick and intuitive operation. Ideal for temperatures around the freezing point and below. For quick, precise change from heating to cooling phases in ramp operation. Low-noise! Temperature range: 0°C to +60°C (ICP 55) Temperature range: -12°C to +60°C (ICP 110 - 750) Max. temperature deviations precision ±0.1°C, uniformity (acc. to DIN) at 37°C ±0.3°C. Turbulence-free internal air-ventilation, separation of the thermal jacket from the interior prevents the occurrence of drying out the loads through the cooling-device. Energy saving heating/cooling system (R134a), highly efficient automatic defrosting system. Segment-dependent control of the internal air-ventilation in 10% steps (from 10 to 100%). 2 stainless steel grids are supplied. Series IPP (SingleDISPLAY) and IPPplus (TwinDISPLAY) Peltier-cooled The cooled incubator generation with the unique ControlCOCKPIT for quick and intuitive operation. With Peltier technology for transition-free heating and cooling within a single system. Environmentally compatible, low energy consumption. Low vibration and nearly noiseless. Temperature range: 0°C to +70°C Max. temperature deviations precision better than 0.1°C, uniformity better than ±0.4°C (at +10 and +37°C) Forced ventilation by Peltier fan. Energy-saving, refrigerant-free Peltier heating/cooling system (heat-pump principle). 1 stainless steel grid each is supplied with the model sizes 30 and 55; on models of sizes 110 to 750 there are 2 stainless steel grids. Series IPS Peltier-cooled with SingleDISPLAY Temperature range: +14°C to +45°C Max. temperature deviations precision better than 0.2°C, uniformity better than ±0.6°C (at +18 and +25°C) Forced ventilation by Peltier fan. Energy-saving, refrigerant-free Peltier heating/cooling system (heat-pump principle). 2 stainless steel grids are comprised in the delivery. SingleDISPLAY: ControlCOCKPIT with one TFT color graphic display - Available parameters on the ControlCOCKPIT: temperature (Celsius or Fahrenheit), fan speed, exhaust air flap position, program runtime, time zones, daylight-saving time
- One temperature sensor Pt100 DIN class A in a 4-wire circuit
- Software AtmoCONTROL for data read out and programming of the data logger via Ethernet interface (90 days test version as free download).
- Ethernet interface on the rear of the appliance for reading out the protocol log
- Double overtemperature protection: Electronic temperature monitoring with freely adjustable monitoring temperature, with option A6 TWW/TWB (protection class 3.1 or 2), mechanical temperature limiter TB acc. to DIN 12 880
TwinDISPLAY: ControlCOCKPIT with two TFT color graphic displays - Available parameters on the ControlCOCKPIT: temperature (Celsius or Fahrenheit), fan speed, exhaust air flap position, program runtime, time zones, daylight-saving time
- Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error
- Software AtmoCONTROL on USB pen drive for programming, managing and transferring of programs via Ethernet interface or USB port
- HeatBALANCE function for application specific adjustment of heating power distribution (balance) between the upper and lower heating groups in an adjustment range between -50% and +50%
- ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function
- Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)
- Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading and implementing programmes and for online logging
- Multiple overtemperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.1 or 2 resp. 3.3) and mechanical temperature limiter TB (protection class 1) acc. to. DIN 12 880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN / MAX values for over/undertemperature
Basic equipment for ICP/IPP/IPPplus/IPS: - PID microprocessor regulation with integrated autodiagnosis system
- Interior made of stainless steel, reinforced by deep drawn ribbing, material no. 1.4301 (ASTM 304), for integraton of the ceramic-and-metal-coated large-area heating
- Structured stainless steel housing, rear of zinc-plated steel
- ControlCOCKPIT for operation and adjustment of all parameters
- High-temperature connectors on the rear of the appliance for single-phase power connection according to country specific systems and IEC standards
- Internal data logger with a storage capacity of at least 10 years
- German, English, French, Spanish language settings available on the ControlCOCKPIT
- Digital timer, adjustable between 1 minute and 99 days
- The SetpointWAIT function guarantees that the process time does not start until the set temperature is reached at all measuring points - optional for temperature values recorded by the freely positionable Pt100 sensors inside the chamber
- Adjustment of three calibration values for temperature and additional appliance specific parameters directly at the ControlCOCKPIT
- By default, all ICP models as well as IPP 750 and IPS 750 come with lockable rollers for transport
Technical data: |
Interior: | made of stainless steel (No. 1.4301/ASTM 304), deep-drawn. |
Housing: | made of textured stainless steel, rear zinc-plated steel. |
Double doors: | inner glass, outer stainless steel |
Electrical supply: | ICP models 230 V (±10%), 50 Hz IPP/IPS models 230 V (±10 %), 50/60 Hz Other voltages to special order. |
Series ICP
| Type | Outer W × H × D mm | Inner W × H × D mm | Vol. approx. l | Wght approx. kg | P rating approx. W | Order No. | |
| ICP55 | 585×1153×514 | 400×400×330 | 53 | 89 | 1200 | 10 26 00551 | |
| ICP110 | 745×1233×584 | 560×480×400 | 108 | 113 | 1200 | 10 26 01101 | |
| ICP260 | 824×1552×684 | 640×800×500 | 256 | 157 | 1200 | 10 26 02602 | |
| ICP450 | 1224×1613×784 | 1040×720×600 | 449 | 217 | 1200 | 10 26 04501 | |
(1) | ICP750 | 1224×1950×784 | 1040×1200×600 | 749 | 249 | 1200 | 10 26 07501 | |
External dimensions: depth without door handle, depth of door handle 56 mm Interior: depth less 33 mm for the fan
Series IPP/IPPplus
| Type | Outer W × H × D mm | Inner W × H × D mm | Vol. approx. l | Wght approx. kg | P rating approx. W | Order No. | |
| IPP30 | 585×704×524 | 400×320×250 | 32 | 42 | 140 | 10 26 00303 | |
| IPP55 | 585×784×604 | 400×400×330 | 53 | 51 | 275 | 10 26 00553 | |
| IPP110 | 745×864×674 | 560×480×400 | 108 | 78 | 650 | 10 26 01103 | |
| IPP260 | 824×1183×774 | 640×800×500 | 256 | 113 | 820 | 10 26 02603 | |
| IPP750 | 1224×1726×874 | 1040×1200×600 | 749 | 234 | 1200 | 10 26 07503 | |
| IPP30plus | 585×704×524 | 400×320×250 | 32 | 42 | 140 | 10 26 00304 | |
| IPP55plus | 585×784×604 | 400×400×330 | 53 | 51 | 275 | 10 26 00554 | |
| IPP110plus | 745×864×674 | 560×480×400 | 108 | 78 | 650 | 10 26 01104 | |
(2) | IPP260plus | 824×1183×774 | 640×800×500 | 256 | 113 | 820 | 10 26 02604 | |
| IPP750plus | 1224×1726×874 | 1040×1200×600 | 749 | 243 | 1200 | 10 26 07504 | |
External dimensions: depth without door handle, depth of door handle 56 mm Interior: depth less 10 mm for the fan - Peltier
Series IPS
| Type | Outer W × H × D mm | Inner W × H × D mm | Vol. approx. l | Wght approx. kg | P rating approx. W | Order No. | |
| IPS260 | 824×1183×774 | 640×800×500 | 256 | 110 | 550 | 10 26 02605 | |
(3) | IPS750 | 1224×1726×874 | 1040×1200×600 | 749 | 206 | 550 | 10 26 07505 | |
External dimensions: depth without door handle, depth of door handle 56 mm Interior: depth less 10 mm for the fan - Peltier
Accessories
Stainless steel grids (standard version: size 30 and 55: 1 piece, size 110 to 750: 2 pieces)
Stainless steel shelves, perforated
Special equipment
| Article | Order No. | |
| IQ document, OQ/PQ checklist | 10 26 03802 | |
| IQ check list with works test data for oven, OQ/PQ check lists as support for validation by customer | |
| IQ/OQ document, PQ checklist (size 30) | 10 26 03803 | |
| IQ/OQ document with works test data for oven, incl. temperature distribution measurement for a freely selectable temperature value. 9 measuring points acc. to DIN 12 880: 2007-05 (other temperature values available on request), PQ check list as support for validation by customer | |
| IQ/OQ document, PQ checklist (from size 55 on) | 10 26 03804 | |
| IQ/OQ document with works test data for oven, incl. temperature distribution measurement for a freely selectable temperature value. 27 measuring points acc. to DIN 12 880: 2007-05 (other temperature values available on request), PQ check list as support for validation by customer | |
| USB ethernet adapter | 10 26 03810 | |
| Ethernet cable, 5 m | 10 26 03811 | |
| for computer interface | |
| USB user ID stick | 10 26 03812 | |
| User authorization license for oven (user ID program) on Memory-Stick for units with TwinDISPLAY | |
| USB stick with documentation software AtmoCONTROL | 10 26 03813 | |
| and operating instructions for units with SingleDISPLAY (comprised in the standard scope of delivery for units with TwinDISPLAY) | |
We reserve the right to make changes in technical specifications. Dimensions subject to confirmation. All performance data are typical mean values for serial devices. Special equipment as entry ports, Pt100 for external temperature recording, a.s.o. upon request.
|
|
|