Main -> Heating units, refrigerating units -> Temperature controllers -> Climatic chambers, test chambers
Climatic Test Chambers, Memmert
Climatic test chamber CTC / Temperature test chamber TTC

Simulate the perfect atmosphere for climatic and temperature tests, in particular acc. to DIN EN 60068-2-1, 2-2 and 2-3. Provide for unlimited flexibility for controlled material and function tests as well as for ageing tests.

Standard scope of delivery
Ventilation and control:

  • high-performance air fan in working chamber
  • depending on operation status, automatic adaption of fan speed resp. manual adjustment from 10 to 100%
  • adaptive, fuzzy-supported, multifunctional, digital microprocessor PID-controller
  • autodiagnostic system with fault indication on temperature and humidity control
  • 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
  • digital 7-day-programme-timer with real time clock, precise minute setting
  • integrated timer for tempering profiles of up to 40 ramps, each segment adjustable from 1 min. up to 999 hours
  • digital display (LED) of all set parameters, such as temperature, weekdays, time, humidity, fan speed and set-up values - language to be chosen in setup
  • digital display of set values (resolution: 0.1°C below 99.9°C, 0.5°C above 100°C) and actual values (resolution: 0.1°C) of temperature (LED)
  • long-term logging (ring store) of all relevant data, GLP-conforming as data logger - 1024 kB
  • programme stored on power failure
  • parallel printer interface (incl. real-time clock with date function) for printing logging files, suitable all PCL3-compatible ink-jet printers (USB available via converter, see accessories)
  • USB interface including MEMMERT Software "Celsius" with added functions for programming and documentation
  • chip card control incl. one MEMoryCard XL with 32 kB (up to 40 ramps)
  • incl. works calibration certificate for -20 °C and +160 °C

Additional standard equipment for CTC:

  • active control for humidifying and dehumidifying (10-98% rh) with digital display of relative humidity - resolution of display: 0,5%, setting accuracy: 1%
  • humidity supply with distilled water from two 10-l- tanks integrated in unit on telescopic slide with automatic change over
  • 2 x 10 l tanks as condensate collector, on telescopic slide
  • incl. works calibration certificate for +30°C at 60% rh

 

Technical data:

Interior:

made of easy-to-clean stainless steel mat. no. 1.4301 (ASTM 304), with

  • 1 stainless steel grid
  • 1 lateral entry port, right, 80 mm diam. with silicone plug
  • Halogen interior lighting 2 x 25 W, switchable

Casing:

Textured stainless steel casing, rear zinc-plated steel

Heating concept:

  • high-performance ring heaters with optimised air circulation
  • door heating to avoid condensate

Cooling concept:

  • refrigerator with twin compressor
  • speed adjustable condenser fan

Multiple overtemperature protection:

  • with audible and visual alarm in case of over-/undertemperature and over-/underhumidity, open door and empty water tank (CTC)
  • independently working, digitally adjustable electronic overtemperature controller TWW, protection class 3.3
  • additional integral over- and undertemperature protection "ASF" (Auto-Safety-Function) automatically following the set value at a preset tolerance range
  • mechanical temperature limiter TB protection class 1 switching the heating off at approx. 10 °C above max. oven temperature

Temperature range:

CTC/TTC without humidity control:

from -42°C up to +190°C

CTC with humidity control:

from +10°C up to +95°C



TypeOuter W × H × D
mm
Inner W × H × D
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
(1)CTC 256898 × 1730 × 1100640 × 670 × 597256297700010 26 02561
TTC 256898 × 1730 × 1100640 × 670 × 597256297700010 26 02562



Outside dimensions: Width plus 20 mm for silicone plug and 5 mm for interfaces, depth without door lock, door handle depth 50 mm

Accessories CTC 256 / TTC 256
ArticleOrder No.
Full-sight glass door, heated10 26 02510
Additional stainless steel grid10 26 02513
External control and logging package10 26 02519
IQ check list10 26 02521

with works test data as support for validation by customer

OQ check list10 26 02522

with works test data for one freely selectable humidity (only CTC) and temperature value incl. temperature distribution survey for 27 measuring points to DIN 12880:2007-05

Oven-related user authorization card41 26 11021

prevents undesired manipulation by unauthorized third parties

RS485 interface instead of USB10 26 04022
Temperature profile write/read unit10 26 04014

for programming via PC, for writing to and reading from the chip card (up to 40 ramps)

Additional chip card, blank, formatted10 26 04015

32 kb MEMoryCard XL for a maximum of 40 ramps




Other special equipment, for example Ethernet interface, additional PT100s etc. are available on request.
We reserve the right to make changes in technical specifications. Dimensions subject to confirmation.
All performance data are typical mean values for serial devices.