Main -> Heating units, refrigerating units -> Temperature controllers -> Climatic chambers, test chambers |
Humidity chambers, Memmert |
Ramp programming for temperature and humidity, active humidity control, and the precise temperature regulation provide for a controlled, physiologically ideal environment for the simulation of real conditions. Suitable for a wide range of applications from building physics over corrosion tests to biological research.
Models HCP
Standard scope of delivery Venting and control systems: - homogenous atmosphere and temperature distribution in the interior by means of encapsulated and turbulence-free ventilation system
- adaptive, fuzzy-supported, multi-functional and digital PID micro-processor control
- self-diagnostics system to find errors for temperature and humidity control
- 2 Pt100-sensor DIN-class A in 4-conductor measurement with changing monitoring and function adaption at exactly the same temperature
- digital real-time week programme-clock with day group functions, can be set by the minute
- relative-time countdown clock for tempering profiles with up to 40 ramps, each segment can be set from 1 min to 999 hrs
- digital display of all temperature, weekday, time, humidity and set-up default values - language selection via set-up
- digital target value specifications (resolution 0.1°C to 99.9°C, 0.5°C from 100°C) and display of current value (resolution 0.1°C) of temperature values (LED)
- active humidification and dehumidification (20-95% rh) with digital display of relative humidity display resolution 0.5%, setting accuracy 1%
- humidification with distilled water from external container by means of self-suction pump
- integrated GLP-conform long-term logging (ring buffer) of all relevant data as data logging - 1024 kB
- saving the programme run in case of power failure
- parallel printer interface (incl. real-time clock with date function) for all PCL3-compatible ink jet printers for GLP-conform documentation
- USB-interface incl. Memmert programming and logging software "Celsius" with advanced features
- chip-card control incl. 1 MEMoryCard XL with 32 kB memory (max. 40 ramps)
- 2nd chip-card (STERICard) for interior sterilisation with fixed values (4 hrs/160°C) without removing sensors
Heating concept: - large-sized multi-function allround heating with additional door and backwall heating to avoid condensate formation
- incl. factory calibration certificate at +60°C
Multiple overheating protection: - with acoustic and optical alarm in case of excess or insufficient temperature and rh, warning when door is open and water container is empty
- separate, electronic temperature selection guard TWW, protection class 3.1, can be set digitally
- additional integrated monitoring of excess or insufficient temperature "ASF", which automatically follows the target value in freely selectable distance
- mechanic temperature limit TB protection class 1 to turn off heating at ca. 10°C above max. chamber temperature
Structure stainless-steel housing: - fully insulated stainless-steel door with double lock and 4-point adjustment
- inner glass door
- fully galvanised stainless-steel back wall
Interior: - easy-to-handle stainless-steel workspace enforced with deep-drawing ribbing W.-St. 1.4301 (ASTM 304), hermetically welded, electropolished
- 2 shelves made of stainless steel, perforated
Temperature range: - with humidity: from +20°C (but min. 8°C above room temperature) to +90°C
- without humidity: von +20°C (but min. 8°C above room temperature) to +160°C
- temperature deviation at 50°C: < ±0.3°C
Power / Power input:
| Type | Outer dimensions W × H × D mm | Inner dimensions W × H × D mm | Vol. approx. l | Wght approx. kg | P rating W | Order No. | |
(1) | HCP 108 | 710×778×550 | 560×480×400 | 108 | 70 | 1000 | 10 26 00108 | |
(2) | HCP 153 | 630×938×650 | 480×640×500 | 153 | 80 | 1500 | 10 26 00153 | |
| HCP 246 | 790×938×750 | 640×640×600 | 246 | 110 | 2000 | 10 26 00246 | |
External dimensions: Depth without door handle, depth of door handle 38 mm. Inner dimensions: Depth is reduced by 25 mm for the fan.
Special equipment
| Article | Order No. | |
| Stainless steel shelves, perforated, for type 108/110/160 | 10 26 04031 | |
| Extra stainless steel shelf, perforatet, for size 153, full width | 10 26 04033 | |
| Extra stainless steel shelf, perforatet, for size 246, full width | 10 26 04034 | |
| Grid for type 108/110/160 | 10 26 05008 | |
| stainless steel grid for HCP 153 | 10 26 05009 | |
| stainless steel grid for HCP 246 | 10 26 05010 | |
| IQ check list | 10 26 02521 | |
| with works test data as support for validation by customer | |
| OQ checklist | 10 26 01023 | |
| with cabinet-related factory data, for one freely selectable humidity and temperature value; includes temperature distribution measurement for 27 measuring points pursuant to DIN 12 880: 2007-05 as support for validation at the customer side | |
| Celsius FDA-Edition | 41 26 11013 | |
| Software conforming to FDA (no domain network support) | |
| Oven-related user authorization card | 41 26 11021 | |
| prevents undesired manipulation by unauthorized third parties | |
| Ethernet interface instead of USB | 10 26 04024 | |
| incl. Software "Celsius Ethernet Edition" | |
| RS232 interface instead of USB | 10 26 04021 | |
| RS485 interface instead of USB | 10 26 04022 | |
| Temperature profile write/read unit | 10 26 04014 | |
| for programming via PC, for writing to and reading from the chip card (up to 40 ramps) | |
| Additional chip card, blank, formatted | 10 26 04015 | |
| 32 kb MEMoryCard XL for a maximum of 40 ramps | |
| USB connection cable for computer interface | 10 26 04023 | |
| Start-up and brief training (Germany only) through Memmert service | 10 26 04012 | |
Special versions like ducts in the rear wall, additional Pt100s etc. are available on request.
We reserve the right to make changes in technical specifications. Dimensions subject to confirmation. All performance data are typical average values for standard appliances.
|
|
|