Main -> Heating units, refrigerating units -> Temperature controllers -> Climatic chambers, test chambers
Constant Climate Chambers, Memmert
Constant climate chambers HPP

The constant climate chamber generation with the unique ControlCOCKPIT for quick, intuitive operation. Perfect for stability tests, climate-controlled storage and conditioning. Highly economical and maintenance-free.
With TwinDISPLAY.

TwinDISPLAY: ControlCOCKPIT with two TFT color graphic displays

  • Available parameters on the ControlCOCKPIT: temperature (Celsius or Fahrenheit), fan speed, exhaust air flap position, program runtime, time zones, daylight-saving time
  • Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error
  • Software AtmoCONTROL on USB pen drive for programming, managing and transferring of programs via Ethernet interface or USB port
  • HeatBALANCE function for application specific adjustment of heating power distribution (balance) between the upper and lower heating groups in an adjustment range between -50% and +50%
  • ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function
  • Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)
  • Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading and implementing programs and for online logging
  • Multiple overtemperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.1 or 2 resp. 3.3) and mechanical temperature limiter TB (protection class 1) acc. to. DIN 12 880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN / MAX values for over/undertemperature

Basic equipment:

  • PID microprocessor regulation with integrated autodiagnosis system
  • Interior made of stainless steel, deep drawn, material no. 1.4301 (ASTM 304)
  • Structured stainless steel housing, rear of zinc-plated steel
  • ControlCOCKPIT for operation and adjustment of all parameters
  • High-temperature connectors on the rear of the appliance for single-phase power connection according to country specific systems and IEC standards
  • Internal data logger with a storage capacity of at least 10 years
  • German, English, French, Spanish language settings available on the ControlCOCKPIT
  • Digital timer, adjustable between 1 minute and 99 days
  • The SetpointWAIT function guarantees that the process time does not start until the set temperature is reached at all measuring points - optional for temperature values recorded by the freely positionable Pt100 sensors inside the chamber
  • Adjustment of three calibration values for temperature and additional appliance specific parameters directly at the ControlCOCKPIT (f.e. relative humidity)

Heating and cooling system:

  • energy-saving Peltier heating/cooling system integrated in back wall

Temperature range:

  • from 0°C to +70°C
  • temperature deviation at 10°C and 37°C: ≤ ± 0,5°C, deviations precision ± 0,2°C

Power:

  • 230 V ± 10%), 50/60 Hz
TypeOuter dimensions W × H × D
mm
Inner dimensions W × H × D
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
HPP110745×864×674560×480×400108 80 65010 26 04110
(1)HPP260824×1183×774640×800×500256122 82010 26 04260
HPP7501224×1726×8741040×1200×600749208120010 26 04750



External dimensions: Depth without door handle, depth of door handle 56 mm.

Inner dimensions: Depth is reduced by 10 mm for the fan - Peltier

Light modules for HPP 110, 260 and 750

10 LED light strips (HPP 110), 14 LED light strips (HPP 260/750) arranged on the side wall of the interior, illumination strength 10,000 LUX), programme-controlled dimming from 0 to 100% in 1 % steps, ramp programming in combination with temperature and humidity

Articlefor sizeOrder No.
Cold white light module 5,500 Kelvin11010 26 01087
Cold white light module 5,500 Kelvin26010 26 01089
Cold white light module 5,500 Kelvin75010 26 01090
Cold white light module 5,500 Kelvin + warm white 2,700 Kelvin11010 26 01088
Cold white light module 5,500 Kelvin + warm white 2,700 Kelvin26010 26 01091
Cold white light module 5,500 Kelvin + warm white 2,700 Kelvin75010 26 01092


Special equipment
ArticleOrder No.
Grid for type 108/110/16010 26 05008
Grid for type 26010 26 01556
Grid for type 450/75010 26 01557
Stainless steel shelves, perforated, for type 108/110/16010 26 04031
Stainless steel shelves, perforated, for type 26010 26 01566
Stainless steel shelves, perforated, for type 450/75010 26 01567
IQ document, OQ/PQ checklist10 26 03802

IQ check list with works test data for oven, OQ/PQ check lists as support for validation by customer

IQ/OQ document, PQ checklist (from size 55 on)10 26 03804

IQ/OQ document with works test data for oven, incl. temperature distribution measurement for a freely selectable temperature value. 27 measuring points acc. to DIN 12 880: 2007-05 (other temperature values available on request), PQ check list as support for validation by customer

IQ/OQ document, PQ checklist10 26 03806

IQ/OQ document with works test data for oven, incl. temperature distribution measurement for a freely selectable temperature and humidity value. 27 measuring points acc. to DIN 12 880: 2007-05, PQ check list as support for validation by customer

IQ/OQ document, PQ checklist10 26 03808

IQ/OQ document with works test data for oven, incl. temperature distribution measurement for a freely selectable temperature and humidity value. 27 measuring points acc. to DIN 12 880: 2007-05, PQ check list as support for validation by customer (for HPP models with light module)

USB ethernet adapter10 26 03810
Ethernet cable, 5 m10 26 03811

for computer interface

USB user ID stick10 26 03812

User authorization license for oven (user ID program) on Memory-Stick for units with TwinDISPLAY




Special versions like lockable door, duct, additional Pt100s etc. are available on request.