Main -> Heating units, refrigerating units -> Temperature controllers -> Climatic chambers, test chambers |
Constant Climate Chambers, Memmert |
Constant climate chambers HPP
The constant climate chamber generation with the unique ControlCOCKPIT for quick, intuitive operation. Perfect for stability tests, climate-controlled storage and conditioning. Highly economical and maintenance-free. With TwinDISPLAY. TwinDISPLAY: ControlCOCKPIT with two TFT color graphic displays - Available parameters on the ControlCOCKPIT: temperature (Celsius or Fahrenheit), fan speed, exhaust air flap position, program runtime, time zones, daylight-saving time
- Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error
- Software AtmoCONTROL on USB pen drive for programming, managing and transferring of programs via Ethernet interface or USB port
- HeatBALANCE function for application specific adjustment of heating power distribution (balance) between the upper and lower heating groups in an adjustment range between -50% and +50%
- ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function
- Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)
- Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading and implementing programs and for online logging
- Multiple overtemperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.1 or 2 resp. 3.3) and mechanical temperature limiter TB (protection class 1) acc. to. DIN 12 880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN / MAX values for over/undertemperature
Basic equipment: - PID microprocessor regulation with integrated autodiagnosis system
- Interior made of stainless steel, deep drawn, material no. 1.4301 (ASTM 304)
- Structured stainless steel housing, rear of zinc-plated steel
- ControlCOCKPIT for operation and adjustment of all parameters
- High-temperature connectors on the rear of the appliance for single-phase power connection according to country specific systems and IEC standards
- Internal data logger with a storage capacity of at least 10 years
- German, English, French, Spanish language settings available on the ControlCOCKPIT
- Digital timer, adjustable between 1 minute and 99 days
- The SetpointWAIT function guarantees that the process time does not start until the set temperature is reached at all measuring points - optional for temperature values recorded by the freely positionable Pt100 sensors inside the chamber
- Adjustment of three calibration values for temperature and additional appliance specific parameters directly at the ControlCOCKPIT (f.e. relative humidity)
Heating and cooling system: - energy-saving Peltier heating/cooling system integrated in back wall
Temperature range: - from 0°C to +70°C
- temperature deviation at 10°C and 37°C: ≤ ± 0,5°C, deviations precision ± 0,2°C
Power:
| Type | Outer dimensions W × H × D mm | Inner dimensions W × H × D mm | Vol. approx. l | Wght approx. kg | P rating approx. W | Order No. | |
| HPP110 | 745×864×674 | 560×480×400 | 108 | 80 | 650 | 10 26 04110 | |
(1) | HPP260 | 824×1183×774 | 640×800×500 | 256 | 122 | 820 | 10 26 04260 | |
| HPP750 | 1224×1726×874 | 1040×1200×600 | 749 | 208 | 1200 | 10 26 04750 | |
External dimensions: Depth without door handle, depth of door handle 56 mm. Inner dimensions: Depth is reduced by 10 mm for the fan - Peltier
Light modules for HPP 110, 260 and 750
10 LED light strips (HPP 110), 14 LED light strips (HPP 260/750) arranged on the side wall of the interior, illumination strength 10,000 LUX), programme-controlled dimming from 0 to 100% in 1 % steps, ramp programming in combination with temperature and humidity
| Article | for size | Order No. | |
| Cold white light module 5,500 Kelvin | 110 | 10 26 01087 | |
| Cold white light module 5,500 Kelvin | 260 | 10 26 01089 | |
| Cold white light module 5,500 Kelvin | 750 | 10 26 01090 | |
| Cold white light module 5,500 Kelvin + warm white 2,700 Kelvin | 110 | 10 26 01088 | |
| Cold white light module 5,500 Kelvin + warm white 2,700 Kelvin | 260 | 10 26 01091 | |
| Cold white light module 5,500 Kelvin + warm white 2,700 Kelvin | 750 | 10 26 01092 | |
Special equipment
| Article | Order No. | |
| Grid for type 108/110/160 | 10 26 05008 | |
| Grid for type 260 | 10 26 01556 | |
| Grid for type 450/750 | 10 26 01557 | |
| Stainless steel shelves, perforated, for type 108/110/160 | 10 26 04031 | |
| Stainless steel shelves, perforated, for type 260 | 10 26 01566 | |
| Stainless steel shelves, perforated, for type 450/750 | 10 26 01567 | |
| IQ document, OQ/PQ checklist | 10 26 03802 | |
| IQ check list with works test data for oven, OQ/PQ check lists as support for validation by customer | |
| IQ/OQ document, PQ checklist (from size 55 on) | 10 26 03804 | |
| IQ/OQ document with works test data for oven, incl. temperature distribution measurement for a freely selectable temperature value. 27 measuring points acc. to DIN 12 880: 2007-05 (other temperature values available on request), PQ check list as support for validation by customer | |
| IQ/OQ document, PQ checklist | 10 26 03806 | |
| IQ/OQ document with works test data for oven, incl. temperature distribution measurement for a freely selectable temperature and humidity value. 27 measuring points acc. to DIN 12 880: 2007-05, PQ check list as support for validation by customer | |
| IQ/OQ document, PQ checklist | 10 26 03808 | |
| IQ/OQ document with works test data for oven, incl. temperature distribution measurement for a freely selectable temperature and humidity value. 27 measuring points acc. to DIN 12 880: 2007-05, PQ check list as support for validation by customer (for HPP models with light module) | |
| USB ethernet adapter | 10 26 03810 | |
| Ethernet cable, 5 m | 10 26 03811 | |
| for computer interface | |
| USB user ID stick | 10 26 03812 | |
| User authorization license for oven (user ID program) on Memory-Stick for units with TwinDISPLAY | |
Special versions like lockable door, duct, additional Pt100s etc. are available on request.
|
|
|