Main -> Volumetry -> Dosing devices -> Dispensers
Dispensette® TA - Bottle-Top Dispenser for Trace Analysis, BRAND
1 Dispensette® TA - bottle-top dispenser for trace analysis

The well-known product range of the bottle-top dispenser Dispensette® provides dispensing task equipment for a great variety of applications.

The Dispensette® TA is ideal for high-precision dispensing of high-purity media for trace analysis. With Pt-Ir valve spring suitable for dispensing hydrofluoric acid.

General properties of the bottle-top dispenser Dispensette®TA

  • Plastics in contact with media consist of high-purity materials such as PTFE, ETFE, FEP, and PFA. The purest sapphire is used for replaceable valves. Depending on the application, platinum-iridium or tantalum are available as spring materials.
  • A field-tested cleaning process before use in trace analysis is described in the operating manual.
  • To avoid contamination of the content of the bottle, we recommend using the unit without the SafetyPrimeTM recirculation valve.
  • GL45 thread and the adapters comprised in the delivery match most commonly used test tubes.
  • Bottle label always visible thanks to 360° rotating valve block
  • Telescopic suction tube can be set for different bottle heights
  • Easy to disassemble for replacement of the dispensing cartridge

Scope of delivery:
Dispensette®, TA, conformity certified, with quality certificate, discharge tube, telescopic filling tube, SafetyPrime™ recirculation valve and recirculation tube (optional), installation wrench and adapter for thread GL 28 / S 28 (ETFE), GL 32 (ETFE) and S40 (PTFE) and instruction manual.

TypeVol.
ml
Grad.
ml
A ≤ ±
%
A ≤ ±
µl
CV ≤
%
CV ≤
µl
Order No.
Dispensette® TA, Pt/Ir spring, without SafetyPrime™ recirculation valve1–100.20.51)501)0.11)101)10 49 74040
Dispensette® TA, Pt/Ir spring, with SafetyPrime™ recirculation valve1–100.20.51)501)0.11)101)10 49 74041
Dispensette® TA, tantalum spring, without SafetyPrime™ recirculation valve1–100.20.51)501)0.11)101)10 49 74240
Dispensette® TA, tantalum spring, with SafetyPrime™ recirculation valve1–100.20.51)501)0.11)101)10 49 74241



1) Adjusted to "Ex". Error limits according DIN EN ISO 8655-5 related to the nominal capacity (= maximum volume) indicated on the instrument, obtained when instrument and distilled water are equilibrated at uniform temperature (20 °C) and with uniform and smooth dispensing. The results are significantly below the error limits specified in DIN EN ISO 8655-6. Conformity certified according to DIN 12 600. R = Correctness, VK = Variation coefficient.

Accessories
ArticleOrder No.
(2)Discharge tube, Pt/Ir valve spring10 49 07955

with integrated valve, drain tube cap PTFE, nominal capacity 10 ml, length 90 mm

Discharge tube, tantalum valve spring10 49 07956

with integrated valve, drain tube cap PTFE, nominal capacity 10 ml, length 90 mm

(3)Dispensing cartridge with safety ring10 49 07542

Nominal volume 10 ml, calibrated, incl. quality certificate

(4)SafetyPrime™ recirculation valve, Pt/Ir valve spring10 49 06086
SafetyPrime™ recirculation valve, tantalum valve spring10 49 06087
(5)Filling valve with seal10 49 06622
(6)Telescopic filling tube, FEP, length 70 – 140 mm10 49 74202
Telescopic filling tube, FEP, length 125 – 240 mm10 49 74203
Telescopic filling tube, FEP, length 195 – 350 mm10 49 74208
Telescopic filling tube, FEP, length 250 – 480 mm10 49 74201