703 laboratory conductivity meter with 4-electrode cell.
With the 703 laboratory conductivity meter you can use either 4-electrode or 2-electrode cells. The measuring input is automatically switched over to the appropriate operating mode. Temperature compensation takes place either automatically with the Pt 100 or Pt 1000 temperature probes or manually.
The standard equipment RS 232 interface allows immediate further processing of the measured values by computer. Also connecting a printer is no problem. Records of the parameter settings, the calibration and the unit diagnostics can be output directly to a printer. This provides you with comprehensive GLP-compatible documentation at the press of a button. The unit automatically selects the display range with the greatest possible resolution. Naturally, the desired display range can also be specified manually. The large, bright LED display allows simultaneous readout of two measured values, e. g. conductivity and temperature. The 14-segment display can show alphanumeric
Technical data: |
External (W×H×D): | 244 × 95 × 255 mm |
Weight: | approx. 2 kg |
Case: | glass-reinforced polyamide 12, rust-proof stainless steel cover, protection type IP 54, prepared for fitting the ZU 6954 attachment stand |
Power supply: | 230 V~ -15% + 10%, 48 to 62 Hz, < 10 VA, optional 115 V~ (opt. 363) |
Display: | alphanumeric, 2 × 4 digits, 14-segment LED, 13 mm high characters, 3 status indicators with Sensoface |
Ranges: |
Cond: | 0,000 to 9,999 µS/cm 0000 to 9999 mS/cm -50.0 to +150.0 |
Measurement error: |
Cond.: | < 0,5% of MV ±2 digit |
°C: | < 0,3 K |
Resolution: | to 0,001µS/cm |
Reproducibility: | < 0,1% of measured value |
Temperature compensation: | -50 to +150°C Pt 100/Pt1000 (autom. selection) or manual, linear TC characteristic 0,00 to 9.99%/K, reference temperature 20°C/25°C selectable |
Permissible cell constant: | 0,001 to 199,9 cm-1, adjustable |
Cell standardization: | operating modes - automatic by cell constant determination with NaCl or KCl solution - direct entry of cell constant |
Monitoring of sensor and meter (GLP): | Sensoface® provides information - for selection of 2-electrode cells - on too great a difference between reference and measuring temperature - for handling of 4-elec trode cells - in case of clock memory loss - in case of irregular checking of measuring equipment |
Unit self-test: | test of measuring electronics including recorder output, segment and keypad test, RAM, EPROM and EEPROM |
GLP records: | parameter settings, calibration, unit diagnostics (DIN ISO 9000), galvanically isolated |
Analog output: |
Cond.: | 1 mV/µS·cm-1 1 mV/mS·cm-1 |
°C: | 10 mV/°C |
Interface: | RS 232 |
Printer control: | for ZU 0244 printer or any other standard printer with serial interface |
EMC directive: | 89 / 366 / EEC |
Standards: | EN 61326 / VDE 0843 Part 20 2002-3 |
4-electrode cell: |
Ranges: | 1 µS/cm to 1 S/cm -20°C to +100°C |
System carrier: | glass |
Ring electrodes (4): | platinum, bright |
Sheath tube: | dia. 16 mm KPG® glass, replaceable |
Immersion depth (min/max): | 60/80 mm |
Temperature probe: | Pt 1000, fast responding |
Time constant: | approx. 1 cm-1 |
Connection cable: | approx. 1 m |
| Article | Order No. | |
| 4-electrodes sensor, body glass (adapter ZU 0290 required) | 61 60 96985 | |
| Calibration solution (1 ampoule for 0.1 mol/l NaCl solution) | 61 60 96945 | |
| Pt 1000 temperature probe ZU 6959 | 61 60 96959 | |
| for temperature measurements with short adjustment time, stainless steel, -10 to +100°C | |
| Spare sheath tube ZU 0180 for cell | 61 60 97180 | |
| Attachment stand ZU 6954 | 61 60 96954 | |
| for accommodating the immersion stirrer and any 3 probes, for connecting directly to the pH meter / conductivity meter | |
| Interface cable ZU 0152 | 61 60 97152 | |
| for connecting to a computer (EMC special cable) | |
| Interface cable ZU 0245 | 61 60 97245 | |
| for connecting to a printer (EMC special cable) | |
| Laboratory printer ZU 0244 | 61 60 97244 | |
| Printing paper ZU 0249 | 61 60 97249 | |
| for ZU 0244 laboratory printer (5 rolls) | |
| Color ribbon ZU 0250 | 61 60 97250 | |
| for ZU 0244 laboratory printer (5 pcs) | |
| Option 363 | 10 60 90363 | |
| auxiliary power 115 V AC | |