Main -> Safety, cleaning, hygiene -> Sterilization -> Steam sterilizers
Steam Sterilizers, SHP Steriltechnik
Steam Sterilizer LABOKLAV 55 – 195

Installation of the unit as either front- or top loader possible. When set up as front loader the units are stackable. Integrated separate steam generator with drying heat elements. Significantly longer service life of heater.
Low in energy consumption due to optimized size of steam generator in relation to chamber volume.

Steam sterilizer fort he universal use in the laboratory, in full compliance with

  • DIN 58950 (pharmaceutical applications)
  • DIN 58951 (laboratory applications)
  • Pressure Device Directive 97/23/EU
  • IEC 61010 – 2 - 43 (Thermo locking device)

LABOKLAV 55 – 195
Basic unit with 55, 80, 100, 135, 160 and 195 Litre chamber volume, Useable as either front- or top loader

  • max. permissible pressure 2,8/-1 bar
  • sterilization temperature 98 °C to 135 °C
  • TÜV tested motor-driven safety quick-action lock with lip seal
  • separate latest-generation steam generator integrated in the unit (4 or 6 kW for Laboklav 100 and up)
  • integrated storage tank for fully demineralized water for steam generation (for automatic or manual filling), level monitoring function
  • thermo-lock pursuant to TRB 402 and DIN 61010-2
  • flexible PT100 reference sensor for fluid sterilization (time triggered via media temperature)
  • gravitation ventilation and pressure change ventilation (waste sterilization)
  • integrated evaporation condensation, water-cooled, temperature-controlled, adjustable
  • validation adapters
  • internal memory chip for storing up to 150 cycles (may be read out later)
  • RS484 interface for data transmission
  • collective connection to drain pipe
  • floor support inside
  • timer function
  • housing made of powder-coated steel. Choose between 3 color combinations: blue/light gray, emerald/light gray, purple/light gray
  • automatic feed water pump
  • microprocessor control with 20 pre-assigned programs
  • individual assignment of all program slots by the user
  • pressure vessel made of SS 316 Ti stainless steel
  • units with full double jacket made of stainless steel from version M or V
  • easy operation
  • code protected access levels
  • safety chains for a safe state of the unit
  • several cycle data can be documented

Fast Cooling of liquids
All units can be supplied with different cooling versions as an option. The cooling time can be reduced dramatically between 50 and 80%. Main feature to achieve short cooling cycles is the double jacket. VE-Water is pumped through the double jacket during cooling, the full contact surface of the chamber ensures a fast heat transfer into the cooling media. No valuable demineralised cooling water is wasted. The cooling water is fed back into the integrated tank and is reused for the next cycle. Automatic cooling of the VE-Water by means of tap water (heat exchange precept).

Vacuum
A powerful watering vacuum pump guarantees a fast and secure deaeration of the sterilisation goods Fractionated or single pre-vacuum programmes are integrated. If ordered with vacuum system vacuum drying is a standard feature. During the drying vacuum hot steam is pulsed through the double jacket to support the drying process.

 

Available variants of LABOKLAV 55 – 195

LAV B:

Basic

LAV M:

integrated fast coolling

LAV MS:

integrated fast recooling with counter pressure

LAV MSL:

integrated fast recooling with counter pressure and additional fan

LAV V:

Vacuum system

LAV MV:

Vacuum system, integrated fast recooling

LAV MSV:

Vacuum system, integrated fast recooling with counter pressure

LAV MSLV:

Vacuum system, integrated fast recooling with counter pressure and additional fan



 

Type

Chamber-
vol.

l

Chamber-
dimensions
(Ø × T)
mm

Installation space
vertical unit D×W×H1)
(Horizontal unit W×H×D)
mm

Loading
capacity2)

Steam
generator

Electr.
requirements

LAV 55

55

410×480

740×765×600
(740×600×765)

approx. 18×1 l Schott

4 kW

3N 400V~16A
(1N 230V~36A)

LAV 80

80

410×610

740×915×600
(740×600×915)

approx. 18×1 l Schott

4 kW

3N 400V~16A
(1N 230V~36A)

LAV 100

100

410×760

740×1065×600
(740×600×1065)

approx. 30×1 l Schott
(3 Lagen)

6 kW

3N 400V~16A
(1N 230V~36A)

LAV 135

135

510×660

840×965×700
(840×700×965)

approx. 30×1 l Schott

6 kW

3N 400V~16A
(1N 230V~36A)

LAV 160

160

510×760

840×1065×700
(840×700×1065)

approx. 45×1 l Schott
(3 Lagen)

6 kW

3N 400V~16A
(1N 230V~36A)

LAV 195

195

510×990

840×1215×700
(840×700×1215)

approx. 45×1 l Schott
(3 Lagen)

6 kW

3N 400V~16A
(1N 230V~36A)

Technical changes alloed
1)Total height, loading height less 90 mm
2)Front loading approx. 15% less



TypeOrder No.TypeOrder No.
LAV 55 B10 16 05510LAV 135 B10 16 01350
LAV 55 M10 16 05511LAV 135 M10 16 01351
LAV 55 MS10 16 05512LAV 135 MS10 16 01352
LAV 55 MSL10 16 05513LAV 135 MSL10 16 01353
LAV 55 V10 16 05514LAV 135 V10 16 01354
LAV 55 MV10 16 05515LAV 135 MV10 16 01355
LAV 55 MSV10 16 05516LAV 135 MSV10 16 01356
LAV 55 MSLV10 16 05517LAV 135 MSLV10 16 01357
LAV 80 B10 16 08010LAV 160 B10 16 01600
LAV 80 M10 16 08011LAV 160 M10 16 01601
LAV 80 MS10 16 08012LAV 160 MS10 16 01602
LAV 80 MSL10 16 08013LAV 160 MSL10 16 01603
LAV 80 V10 16 08014LAV 160 V10 16 01604
LAV 80 MV10 16 08015LAV 160 MV10 16 01605
LAV 80 MSV10 16 08016LAV 160 MSV10 16 01606
LAV 80 MSLV10 16 08017LAV 160 MSLV10 16 01607
LAV 100 B10 16 01000LAV 195 B10 16 01950
LAV 100 M10 16 01001LAV 195 M10 16 01951
LAV 100 MS10 16 01002LAV 195 MS10 16 01952
LAV 100 MSL10 16 01003LAV 195 MSL10 16 01953
LAV 100 V10 16 01004LAV 195 V10 16 01954
LAV 100 MV10 16 01005LAV 195 MV10 16 01955
LAV 100 MSV10 16 01006LAV 195 MSV10 16 01956
LAV 100 MSLV10 16 01007LAV 195 MSLV10 16 01957


Accessories
ArticleOrder No.
Printer for LABOKLAV 55 – 19510 16 02030

Integrated printer for the documentation of all relevant prozcess parameters, real paper printer, stored data can be printed out at a later stage

Software "Documentator"10 16 03030

Software for the visualisation, printing and storage of all relevant process data

Exhaust air filtration10 16 02010

for safety laboratories BSL 2 and 3

Loading system for LABOKLAV 135/16010 16 00160

Comprises the trolley and the loading slide

Loading system for LABOKLAV 19510 16 00195

Comprises the trolley and the loading slide

Lifting device for loading and unloading the LABOKLAV 135 - 19510 16 02040

with swivel arm, for loading and unloading heavy material




Buckets an baskets made of stainless steel on request