Installation of the unit as either front- or top loader possible. When set up as front loader the units are stackable. Integrated separate steam generator with drying heat elements. Significantly longer service life of heater.
Low in energy consumption due to optimized size of steam generator in relation to chamber volume.
Steam sterilizer fort he universal use in the laboratory, in full compliance with
- DIN 58950 (pharmaceutical applications)
- DIN 58951 (laboratory applications)
- Pressure Device Directive 97/23/EU
- IEC 61010 – 2 - 43 (Thermo locking device)
LABOKLAV 55 – 195
Basic unit with 55, 80, 100, 135, 160 and 195 Litre chamber volume, Useable as either front- or top loader
- max. permissible pressure 2,8/-1 bar
- sterilization temperature 98 °C to 135 °C
- TÜV tested motor-driven safety quick-action lock with lip seal
- separate latest-generation steam generator integrated in the unit (4 or 6 kW for Laboklav 100 and up)
- integrated storage tank for fully demineralized water for steam generation (for automatic or manual filling), level monitoring function
- thermo-lock pursuant to TRB 402 and DIN 61010-2
- flexible PT100 reference sensor for fluid sterilization (time triggered via media temperature)
- gravitation ventilation and pressure change ventilation (waste sterilization)
- integrated evaporation condensation, water-cooled, temperature-controlled, adjustable
- validation adapters
- internal memory chip for storing up to 150 cycles (may be read out later)
- RS484 interface for data transmission
- collective connection to drain pipe
- floor support inside
- timer function
- housing made of powder-coated steel. Choose between 3 color combinations: blue/light gray, emerald/light gray, purple/light gray
- automatic feed water pump
- microprocessor control with 20 pre-assigned programs
- individual assignment of all program slots by the user
- pressure vessel made of SS 316 Ti stainless steel
- units with full double jacket made of stainless steel from version M or V
- easy operation
- code protected access levels
- safety chains for a safe state of the unit
- several cycle data can be documented
Fast Cooling of liquids
All units can be supplied with different cooling versions as an option. The cooling time can be reduced dramatically between 50 and 80%. Main feature to achieve short cooling cycles is the double jacket. VE-Water is pumped through the double jacket during cooling, the full contact surface of the chamber ensures a fast heat transfer into the cooling media. No valuable demineralised cooling water is wasted. The cooling water is fed back into the integrated tank and is reused for the next cycle. Automatic cooling of the VE-Water by means of tap water (heat exchange precept).
Vacuum
A powerful watering vacuum pump guarantees a fast and secure deaeration of the sterilisation goods Fractionated or single pre-vacuum programmes are integrated. If ordered with vacuum system vacuum drying is a standard feature. During the drying vacuum hot steam is pulsed through the double jacket to support the drying process.
| Article | Order No. | |
| Printer for LABOKLAV 55 – 195 | 10 16 02030 | |
| Integrated printer for the documentation of all relevant prozcess parameters, real paper printer, stored data can be printed out at a later stage | |
| Software "Documentator" | 10 16 03030 | |
| Software for the visualisation, printing and storage of all relevant process data | |
| Exhaust air filtration | 10 16 02010 | |
| for safety laboratories BSL 2 and 3 | |
| Loading system for LABOKLAV 135/160 | 10 16 00160 | |
| Comprises the trolley and the loading slide | |
| Loading system for LABOKLAV 195 | 10 16 00195 | |
| Comprises the trolley and the loading slide | |
| Lifting device for loading and unloading the LABOKLAV 135 - 195 | 10 16 02040 | |
| with swivel arm, for loading and unloading heavy material | |
Buckets an baskets made of stainless steel on request