Main -> Special analysis -> Water, environment -> BOD determination |
BOD AL606 Measurement System, AQUALYTIC |
BOD AL606 Measurement System
The AQUALYTIC® sensor system BOD AL606 allows precise measurement of BD based on the manometric principle. Manometric respirometers relate wxygen uptake to the change in pressure caused by oxygen consumption while maintaining a constant volume. Thanks to the modern integral pressure sensors, it is no longer necessary to use mercury for pressure measurement. The measured values are read out comfortably via the front panel. There are no batteries required in the sensor heads. Advantages - Accurate and direct display of BOD values in mg/l
- User-friendly handling
- User-selectable measuring period from 1 to 28 days (BOD 5, BOD 7, OECD …)
- Measuring ranges from 0 – 40 mg/l to 0 – 4000 mg/l BOD, sample volume related
- Automatic start function following temperature equalisation of the samples
- Mercury-free, environmentally friendly BOD determination
- Inductive stirring system with automatic re-centering of stirring rods
- Direct sample selection
- Interface RS 232
Inductive stirring system The microprocessor-controlled AQUALYTIC® inductive stirring system is non-wearing an maintenance-free. At regular intervals, the magnetic stirring rods are accelerated and slowed down again, taking them up to maximum speed and back down again. This ensures the centralization of the stirring rods. Stirring rods that move away from the center of the bottle are re-centered quickly and reliably.
Technical data: |
Meas. principle: | Manometric; mercury-free; electronic pressure sensor |
Ranges (mg/l O2): | 0 – 40, 0 – 80, 0 – 200, 0 – 400, 0 – 800, 0 – 2000, 0 – 4000 mg/l |
Accuracy: | 0,5 % full scale at 20°C |
Applications: | BOD5, BOD7, OECD 301F |
Result display: | BOD (mg/l), 4 digits, 7-segment LED |
Measurement parameter display: | BOD-range, volume, duration, time of measurement |
Measurement period: | User-selectable, between 1 and 28 days |
Auto result storer: | Up to 28 results, Depending on measurement period |
Storage interval: | houly (1 day) every 2 hours (2 days) daily (3 – 28 days) |
Automatic start funciton: | After temperature equalisation of samples; Can be switched off |
Power supply: | 3 alkaline-manganese batteries ("Baby" cells/size "C") |
Battery life: | 1 year (normal use as BOD5 meter-max. one reading a day); early warning before battery fails |
Clock: | Real-time clock |
Protection class: | IP 54 (sensor head) |
Housing: | ABS |
Approval: | CE |
| Article | Order No. | |
(1) | AQUALYTIC® COD AL606, 6 sensor | 80 77 44406 | |
| complete with 6 sensors and control unit with batteries, inductive stirring unit with power supply, 6 sample bottles, 6 rubber gaskets, 6 stirring rods, 1 overflow flask 157 ml, 1 overflow flask 428 ml, 50 ml potassium hydroxide solution 45%, 50 ml nitrificiation inhibitor solution, instruction manual | |
| AQUALYTIC® COD AL612, 12 sensors | 80 77 44410 | |
| as above but with 12 sensor heads | |
2 Accessories
| Article | Order No. | |
| System check | 10 77 18328 | |
| System check set, BOD CM test tablets, box with 8 tablets | |
| Potassium hydroxide solution 45%, 50 ml | 10 77 18634 | |
| Nitrification inhibitor, 50 ml | 10 77 18642 | |
| BOD bottle for sensors, brown glass, 500 ml | 10 77 18644 | |
| Rubber gasket | 10 77 18636 | |
| Stirring rod | 10 77 18637 | |
| Complete set Overflow flasks | 10 77 18654 | |
| 1 piece of 21,7 ml, 56 ml, 94 ml, 157 ml, 244 ml, 360 ml, 428 ml | |
|
|
|